Programmable Temperature & Humidity Tester
GT-C52 environmental test chamber is essential for evaluating the performance and durability of safety gloves and protective materials under extreme environmental conditions. This material testing machine supports sample conditioning prior to mechanical testing to ensure results comply with international standards.
Specifications| Capacity | 80L | 150L | 225L | 408L | 612L | 1000L |
| Inside dimensions WxHxD(cm) |
40x50x40 | 50x60x50 | 50x75x60 | 60x85x80 | 80x95x85 | 100X100X100 |
| Outside dimensions WxHxD(cm) |
95x136x94 | 100X146X104 | 100x161X117 | 110x171x137 | 130x181x137 | 152 x 210 x 127 |
| Temperature range (optional, can be custom-made) | -20~150℃, | -40~150℃, | -60~150℃ |
| Humidity range | 20%~98%, can be set. | ||
| |||
| Rate of temperature increasing | Room temperature~150℃ | About 45 minutes(non-linear no-load, about 3.5℃/minute | |
| Room temperature~-20℃ | About 45 minutes(non-linear no-load, about 1.2℃/minute) | ||
| Machine precision | Control accuracy (test precision) | Temperature | +/-0.5℃ |
| Humidity | +/-1%RH | ||
| Distribution accuracy (uniformity) | Temperature | +/-1.0℃ | |
| Humidity | +/-2%RH | ||
| Environment temperature | 5℃-28℃ | ||
| Environment relative humidity | ≤85%RH | ||
| Power voltage | 220V AC(+/-10%), protect ground wire, ground resistance≤4Ω. | ||
| Frequency | 50+/-0.5Hz | ||
| Power switch | Air-break switch(40A, knife switch or outlet is prohibited to use.) | ||
| Power | 4.5KW | ||
| Humidification water | Resistivity(≥500MΩ) | ||
| Requirement of ambient air quality | Do not contain high concentration dust and flammable gas, without strong electromagnetic radiation source nearby. | ||
| Power supply | AC1¢3W 380V 50/60HZ, voltage change allowed value: ±10% of fixed voltage. | ||
| Max loading capacity | 8KVA | ||
| Max operating current | 30A | ||