GT- RA16 Dry State Lint Tester

Application
This tester is used for testing the coefficient of linting and other particles generation of surgical drapes, gowns and clean air suits in the drying state.

Standards
• ISO 9073-10
• INDA IST 160.1
• DIN EN 13795-2
• YY/T 0506.4

Technical parameters
1.Diameter of the circular plate:Φ82.8㎜
2.Movement frequency: 60 times/ min
3.Rotating Angle: 180°/time (clockwise & anticlockwise alternating)
4.Max starting distance of 2 circular plates: (188±2)㎜
5.Straight stroke of the moving plates: (120±2)㎜
6.Size of anti-static electricity flexing chamber: 300mm×300mm×300mm
7.Diameter of air collector:(40±5)㎜
8.Power Supply: AC 220V 50Hz
9.Laser dust particle counter

a).Range of particle sizes:0.3μm~25μm
b).8 test channel:0.3μm, 0.5μm, 0.7μm, 1.0μm, 2.0μm, 3.0μm, 5.0μm, 10.0μm
c).Air flow speed:(28.3±1.4)L/min
d).Sampling period:1~59min59s optional
e).Printing function: Built-in printer, can print counting result
f).Maximum sampling concentration: 35,000 particles/L